The Cary 6000i UV-Vis-NIR spectrophotometer has been tailored for unmatched photometric performance in the short-wave infrared (SWIR) 800–1800 nm range as well as the UV-visible 175 to 800 nm.
The superior sensitivity of the narrow-band InGaAs detector is optimally harnessed by preferencing the spectrophotometer’s optical throughput in the SWIR region. Its 600 lines-per-mm NIR diffraction grating, blazed at 1000 nm, concentrates optical dispersion into the 800–1800 nm region, eliminating wasted photons that would otherwise fall beyond the range of the detector. The Cary 6000i is your ultimate tool for materials characterization spectroscopy.
Depth | 710 mm |
Height | 380 mm |
Light Source | Tungsten halogen visible and deuterium arc UV |
Maximum Scanning Speed | UV-Vis 2,000 nm/min NIR 8,000 nm/min |
Photometric System | Double beam |
Power Consumption | Mains supply of 100 – 240 volts AC and frequency 50 – 60 Hz |
Spectral Bandwidth | UV-Vis 0.01 – 5.00 nm NIR 0.04 – 20 nm |
Wavelength | 175 – 3300 nm |
Width | 1020 mm |
Z-height | 20 mm |