Collect full spectral data for postacquisition analysis with the Dionex UltiMate 3000 Photodiode Array Detector. It supports both HPLC and UHPLC applications with eight channels of single-wavelength data acquisition, a 1024-element diode array, and the excellent noise and drift performance of a forward-optic detector. Diode array detection allows full spectrum acquisition at speeds up to 200Hz. Comprehensive three-dimensional data tools in Dionex Chromeleon Chromatography Data System software makes it easy to process and report your results.
Specifications:
Optical System | Single-beam, reverse-optics design with concave holographic grating |
Photodiode Array | 1024-element |
Lamps | Deuterium lamp (30 W) for ultraviolet spectrum analysis Tungsten lamp (15 W) for visible spectrum analysis |
Spectral Resolution | 1 nm |
Pixel Resolution | 0.7 nm |
Wavelength Range | 190 to 800 nm |
Wavelength Accuracy | ±1 nm, self-calibration with deuterium lines, verification with built-in holmium oxide filter |
Noise with Analytical or Semipreparative Cell Installed | <± 10 µAU at 254 nm, 2-second rise time, 4 nm; bandwidth and flowing water at 1.0 mL/min <± 15 µAU at 520 nm, 2-second rise time, 10 nm; bandwidth and flowing water at 1.0 mL/min |
Drift | <1 mAU/h at 254 nm and 520 nm, and flowing water (after warm-up); typically < 500 µAU/h |
Absorbance Range | ?2.7 (wavelength dependent) to 3.5 AU |
Linearity | Corr. Coeff. > 99.95 and < 3% RSD up to 1.5 AU; Typical Corr. Coeff. > 99.99 and < 2.5% RSD up to 1.8 AU |
Power Requirements | 85 to 265 VAC, 47 to 63 Hz auto-sensing; no adjustment necessary); Typical input power 100 W; Maximum line draw 3.5 A at 110 VAC at power-up |
Dimensions (h × w × d) | 19 × 42 × 51 cm |
Control | All parameters and functions controlled by Chromeleon software; Chromeleon PC and detector connected via USB (Universal Serial Bus); 4 LEDs (Power, Run, Deuterium, Tungsten) for status monitoring |
Weight | 19 kg |
Decibel Level | <50 dbA during routine operation 56 dbA at power-up |
This Dionex Ultimate 3000 Photodiode Array Detector PDA-3000 is in good condition. It has been tested here at Spectralab Scientific. A 90-day warranty is included.