Specifications |
|
Wavelength Range |
200–595 nm, programmable |
Noise |
< ± 5 µAU (dry cell, 254 nm,8 nm, 1 s) |
Drift |
< 500 µAU/h |
Wavelength Accuracy |
± 0.75 nm (UV), ± 1.5 nm (Vis) |
Bandwidth |
1.9–400 nm, programmable |
Linearity |
≤1% at 1.5 AU |
Measuring Range |
0–2 AU |
Time Constant |
0.1–4.0 s, programmable |
Power-up Wavelength Verification |
Automatic via internal holmiumoxide filter |
Wavelength Calibration |
Automatic or on demand; over the entire wavelength range |
Optical Bench |
Photodiode array bench withoutmoving parts |
Pixel Resolution |
1.9 nm (UV), 3.3 nm (Vis) |
Lamp |
Deuterium Lamp |
Standard Cell |
10 µL, 9 mm, 10 MPa (1450 psi) |
Biocompatible Cell (optional) |
10 µL, 9 mm, 10 MPa (1450 psi) |
Preparative Cell (optional) |
6 µL, 2 mm, 10 MPa (1450 psi) |
Micro Capillary Cell (optional) |
0.14 µL, 8 mm, 40 MPa (5800 psi) |
This refurbished Dionex UVD 170U UV/VISDetector is in great condition. It has been tested at SpectraLabScientific.
A 90 day warranty is included.