Detection Type |
Double beam photometer |
Signals |
Up to 3 (MWD) or 5 (DAD), each defined by a sample wavelength and bandwidth and an optional reference wavelength and bandwidth. |
Signal Combinations |
Sum of two signals; difference of two signals; ratio of two signals, with definable range and threshold; window plot with definable ratio, range and threshold. |
Noise |
< ±2.0 x 10-5 AU peak-to-peak, at 254 nm with 4 nm bandwidth, flowing water at 1 ml/min, 1 second response time (10-90%). |
Drift |
< 2 x 10-3 AU/hour, at 254 nm, after warm-up |
Wavelength Range |
190-600 nm, selectable in 1 nm increments |
Bandwidth Range |
2-400 nm, selectable in 1 nm increments. |
Wavelength Accuracy |
±1 nm. |
Linear Absorbance Range |
Better than 1% up to 1.5 AU, using acetone at 265 nm. |
Response Time |
8 choices, ranging from 0.1 to 20 seconds (10-90%). |
Spectra |
Storage of at least 8 spectra, with definable wavelength range from 190-600 nm (total number of spectra depends on range defined). Post-run plotting of original or subtracted spectra possible |
Spectra Resolution |
Depends on slit width. |
Absorbance Resolution |
< 5 x 10-6 AU |
Spectra Acquisition |
12.5 ms from 190 to 600 nm; according to the response time settings scans are accumulated for one data point for improved sensitivity. |
Light Source |
Deuterium lamp. |
Flow Cell (Series II) |
SST cell, 8 µl volume with 6 mm path length or 13 µl volume with 10 mm path length and 120 bar (1760 psi) maximum pressure. Optional high-pressure cells (400 bar). |
Display |
2 line by 16 character fluorescent display with real-time display of operating parameters or absorbance and wavelength of maximum absorbance. |
Control |
Integrated keyboard with function keys; parameter editing during run possible; keyboard lock. External control via GPIB with ChemStation or RS-232.Integrated keyboard with function keys; limited functionality and control on DAD. |
Parameters |
Signal definitions (wavelengths/bandwidths), attenuation, response time, zero offset, balance; spectra definitions; external contacts; analog outputs. |
Time-programmable Parameters |
Wavelengths/bandwidths, attenuation, spectra acquisition, external contacts |
Methods |
Battery-backed storage of up to ten methods (depends on length of method), including initial and time-programmed parameters. Automatic startup and shutdown methods. Editing of stored methods possible during run. |
Analog Outputs |
Two analog outputs are available (on DAD optional) for output of signals and/or signal combinations; both 1 V/2 AU or 100 mV/2 AU, user selectable. |
Communications |
START (in- and output), STOP (in- and output), READY (output), SHUTDOWN (output) for synchronization with other LC modules; two external contacts including: 1 relay with 24 V and 250 mA; 1 contact closure with maximum 30 V and 250 mA. |
Safety Aids |
Extensive diagnostics, error detection and display via front-panel LED’s and status logbook. User-definable shutdown method activated in case of error. Leak detection, safe leak handling, leak output signal to shutdown the pump |
Environment |
Temperature range: 5 to 55 °C Humidity: < 85% (non-condensing) |
Power Requirements |
Line voltage: 100-120 or 220-240 VAC ±10%Line frequency: 48-66 Hz Power consumption: 150 VA max |
This refurbished HP 1050 MWD is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.