This Agilent 1100 HPLC G1329A Autosampler is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.
Specifications:
GLP Features | Early maintenance feedback (EMF), electronic records of maintenance and errors |
Communications | Controller-area network (CAN). GPIB (IEEE-448), RS232C, APG-remote standard, optional four external contact closures and BCD vial number output |
Safety Features | Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
Injection Range |
0.1 – 100 µl in 0.1 µl increments Up to 1500 µl with multiple draw (hardware modification required) |
Replicate Injections | 1 – 99 from one vial |
Precision |
Typically < 0.5 % RSD of peak areas from 5 – 100 µl Typically < 1 % RSD of peak areas from 1 – 5 µl |
Minimum Sample Volume | 1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in 300 µl microvial |
Carryover | Typically < 0.1 %, < 0.05 % with external needle cleaning |
Sample Viscosity Range | 0.2 – 50 cp |
Replicate Injections Per Vial | 1 – 99 |
Sample Capacity |
100 × 2-ml vials in 1 tray 40 × 2-ml vials in ½ tray 15 × 6-ml vials in ½ tray (Agilent vials only) |
Injection Cycle Time | Typically 50 s depending on draw speed and injection volume |