This refurbished HP / Agilent 1100 Series G1367A AutoSampler with G2254A Auto Interface is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.
Specification of G1367A Autosampler
Type | Specification |
GLP features | Early maintenance feedback (EMF), electronic records of maintenance and errors |
Communications | Communications Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output |
Safety features | Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
Injection range | 0.1 – 100 µl in 0.1 µl increments Up to 1500 µl with multiple draw (hardware modification required) |
Precision | Typically < 0.5% RSD from 5 – 100 µl, Typically < 1% RSD from 1 – 5 µl variable volume |
Sample viscosity range | 0.2 – 5 cp |
Sample capacity |
2 × well plates (MTP) + 10 × 2 ml vials 100 x 2-ml in one tray 40 x 2-ml in half tray |
Injection cycle time | Typically < 30 s using following standard conditions: Default draw speed: 200 µl/min Default eject speed: 200 µl/min Injection volume: 5 µl |
Carry-over |
Typically < 0.01% using the following conditions: Column: 125 x 4 mm Hypersil ODS, 5 µm Mobile phase: Water/Acetonitrile = 80/20 Flow rate: 1 ml/min Injection volume: 1 µl caffeine (1 mg/ml), 5 µl water to test carryover Outside wash of needle before injection: 20 sec with water using flush port |
Specification of G2254A Auto Interface
Operating Temperature | 5°C to 55°C |
Operating Humidity | < 95%, at 25 – 40°C, non-condensing |
Altitude | Up to 2000M |
Line Voltage | CAN DC in, 24V |
Interfaces | CAN to any Agilent 1100 module |