This refurbished HP / Agilent 1100 Series G1377A μ-WPS AutoSampler is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.
Type | Specification |
GLP features | Early maintenance feedback (EMF), electronic records of maintenance and errors |
Communications | Communications Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output |
Safety features | Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
Injection range | 0.01 – 8 µl in 0.01 µl increments with the small loop capillary 0.01 – 40 µl in 0.01 µl increments with the extended loop capillary |
Precision | Typically < 0.5 % RSD of peak areas from 5 – 40 µl, Typically < 1 % RSD from 1 – 5 µl Typically < 3 % RSD from 0.2 – 1 µl |
Sample viscosity range | 0.2 – 5 cp |
Sample capacity | 2 × well-plates (MTP) + 10 × 2 ml vials 100 x 2 ml in one tray 40 x 2 ml in half tray |
Injection cycle time | Typically < 30 s using following standard conditions: Default draw speed: 4 µl/min Default eject speed: 10 µl/min Injection volume: 0.1 µl |
Carry-over | Typically < 0.05 % using the following conditions: Column: 150 x 0.5 mm Hypersil ODS, 3 µm Mobile phase: Water/Acetonitrile = 85/15 Column Flow rate: 13 µl/min Injection volume: 1 µl caffeine (=25ng caffeine), 1 µl water to test carryover Outside wash of needle before injection: 20 sec with water using flush port |