Type |
Specification |
Sample capacity |
100 x 2 ml vials in 1 tray. Microvials (100 or 300 µl) with sleeves (reduced cooling performance with microvials) |
Settable injection volume |
0.01 to 8 µl with small loop capillary 0.01 to 40 µl with extended loop capillary |
Precision |
Typically < 0.5 % RSD from 5 to 40 µl Typically < 1 % RSD from 1 to 5 µl Typically < 3 % RSD from 0.2 to 1 µl |
Minimum sample volume |
1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in 300 µl microvials |
Carryover |
Typically < 0.1 % without automated needle wash injection volume |
Sample viscosity range |
0.2 – 5 cp |
Recommended pH-range |
1.0 – 8.5 solvent with pH < 2.3 should not contain acids which attack stainless steel. Upper pH range is limited by fused silica capillaries |
Material in contact with solvent |
Stainless steel, sapphire, PTFE, PEEK, fused silica, Vespel |
GLP features |
Early maintenance feedback (EMF), electronic records of maintenance and errors |
Communications |
Controller-area network (CAN). GPIB (IEEE-448), RS232C, APG-remote standard, optional four external contact closures and BCD vial number output |
Safety features |
Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
Housing |
All material recyclable |
This refurbished HP / Agilent 1100 Series G1389A MicroALS is in great condition. It has been tested at SpectraLab Scientific.
Warranty is included.
We have special expertise on analytical equipment.
Multiple available