System includes:
- CBM-40 System Controller
- DGU-405 Degassing Unit
- LC-40B XR Solvent Delivery Module
- SPD-M40 PhotoDiode Array Detector
- CTO-40S Column Oven
- SIL-40C XR AutoSampler
- PC
The Shimadzu Nexera XR System represents a leap forward in HPLC technology, offering advanced analytical performance tailored to modern laboratory needs. Beyond instrument precision, the Nexera XR focuses on lab-wide efficiency by integrating AI-driven capabilities for automatic issue detection and resolution.
With IoT-enabled device networking, the system streamlines lab management, making it easy to monitor instrument status and optimize resource allocation. The Nexera XR also incorporates Shimadzu’s Nexflow technology*, designed to minimize extra-column dispersion while maintaining optimal UHPLC performance for high-speed, high-separation analysis. Backed by over 40 years of expertise in LC innovation, the Nexera XR sets a new benchmark in HPLC systems.
* This is optional for the Shimadzu Nexera system. The effectiveness differs depending on the analysis conditions.
Pressure | 70MPa |
Approx. particle size (*) | 2μm-3μm |
I.D. (*) | 2-4.6mm |
Dual pump | ✔ |
Multi-rinse | Optional |
Key features | This UHPLC-like model is the new standard for the Shimadzu LC series. It is compatible with a wide range of analysis conditions from conventional HPLC analysis to ultra-high separation analysis using core-shell columns with a particle size of 3 μm and below. |
*To achieve maximum functionality of both instruments and columns, we recommend using columns with the particle size and dimensions suggested. However, as long as the maximum pressure of the instrument and columns are not exceeded, other columns can be used.
Piping adjustments may be necessary depending on the analysis conditions. Please contact a Shimadzu sales representative for details.
This UHPLC-like model is the new standard for the Shimadzu LC series. It is compatible with a wide range of analysis conditions from conventional HPLC analysis to ultra-high separation analysis using core-shell columns with a particle size of 3 μm and below.
Nexera XR High Pressure Gradient
Model Name | Features | |
System controller | SCL-40 | Color screen with touch panel and system controller with reservoir tray Number of connectable units : 8 (Using option: 12) Connectable unit : Solvent delivery unit: max. 4, Autosampler: 1, Column oven: max. 4, Detector: max 2, etc. |
Solvent Delivery Pump | LC-40B XR | Pumping method : Parallel-type double plunger (approx. 10 μL/1 stroke) Allowable maximum pressure : 70 MPa Flow rate settings range : 0.0001 – 3.0000 mL/min (1.0 – 70 MPa) 3.0001 – 5.0000 mL/min (1.0 – 44 MPa) 5.0001 – 10.0000 mL/min (1.0 – 22 MPa) Flow rate accuracy : ≦± 1% or ± 2 μL/min, whichever greater (under specified conditions) Gradient mode : High-pressure gradient (2 solvents (LC-40B XR standard) or 3 solvents) Quaternary low-pressure gradient (Only available for LC-40D XR) |
Gradient Mixer | ||
Degassing Unit | DGU-403 | Number of degassed solvents : 3 Degassed flow line capacity : 400 μL/1 line Operating temperature range : 4 to 35°C |
Autosampler | SIL-40C XR | Injection method : Total-volume Injection (standard), loop injection (optional) Allowable maximum pressure : 80 MPa Injection volume : 0.1 to 50 μL, 0.1 to 2000 μL (optional) Injection volume accuracy : ≦± 1% (5 μL injection, n = 20) Injection cycle time : ≦6.7 seconds (under specified conditions) |
CTO-40S | Temperature control type : Forced air circulation Cooling Method : Electronic cooling Temperature control range : Room temperature − 10°C to 85°C Temperature accuracy : ± 0.8°C Temperature precision : ± 0.1°C |
|
UV-VIS Detector | SPD-M40 | Light source :Deuterium (D2) lamp, Tungsten lamp Number of diode elements : 1024 Wavelength range : 190 to 800 nm Wavelength accuracy : ≦± 1 nm Wavelength reproducibility : ≦± 0.1 nm |
Software | Labsolutions LC/GC Multi LC-PDA |