Dimensions and Weights
- Size: 53 cm (h) x 66 cm (w) x 56 cm (d)
- Weight: 43 kg
Environmental Conditions
- Operating temperatures: 10°C to 40°C
- Operating humidity (relative): 5% to 95%
- Line voltage requirements: 101V, 120V, 230V (±10% nominal)
Column Oven
- Dimensions: 28 cm (w) x 20 cm (d) x 28 cm (h)
- Temperature range: ambient +4 °C to 450 °C
- With liquid N2: -99 °C to 450 °C
- With liquid CO2: -65 °C to 450 °C
- Temperature program ramps/holds: 7/8
- Maximum temperature ramp rate: 100 °C/min for all voltages
- Cool down rate: 400 °C to 50 °C in 4.5minutes
- Temperature set-point resolution: 1 °C
Injector Options
- Maximum injectors: three operating concurrently
- Pneumatics: manual or electronic flow control (EFC)
- Columns: capillary columns standard with optional packed column adapter
- Injectors:
- CP-1177Split/Splitless
- 1079 Programmable Temperature Vaporizing (PTV)
- 1061 Flash Megabore/Packed
- 1041 On-Column Megabore/Packed
CP-1177 Split/Splitless
- Pressure range: 0-100 psi
- Total flow: 500 rrillrnin at 10 psi
- Maximum temperature: 450 °C
- Split range: 1-10,000 (column dependent)
1079 Programmable Temperature Vaporizing (PTV)
- Pressure range: 0-100 psi
- Total flow: 500 mL/min at 10 psi
- Temperature range
- 99 °C to 450°C liquid N2
- 65 °C to 450°C liquid CO2
- Maximum temperature ramp rate: 180 °C/min
- Temperature ramps: 4
- Split range: 1-10,000 (column dependent)
- Operational modes
- Temperature ramped splitless
- Coldon-column
- Large Volurne Injection
- Split and splitless
- ChromatoProbe solid sample introduction: optional
1061 Flash Megabore/Packed
Pressure range: 0-100 psi
Total Flow: 50 mL/min (Type 3 EFC)
500mL/min (Type 4 EFC)
Maximum temperature: 450 °C
1041 On-Column Megabore/Packed
Pressure range: 0-100 psi
Total Flow: 50 mL/min (Type 3 EFC)
500mL/min (Type 4 EFC)
Maximum temperature: 450 °C
This refurbished Varian GC CP 3800 is in a great shape. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.