Whether performing routine UV-based applications or pushing the limits of detection with low-level impurity analyses, the Waters 2489 UV/Visible (UV/Vis) Detector is the best detection choice for performance, reliability, and usability. The 2489 UV/Vis Detector is the most sensitive and versatile dual-wavelength absorbance detector available for HPLC. The design combines a high energy Deuterium source, a precise optical design, and low noise, high speed electronics. These advanced capabilities take UV/Vis detection performance to a new level.
Features:
- Enhanced sensitivity via patented TaperSlit™ flow cell, which channels light through the cell for better energy throughput while minimizing RI effects
- Low noise performance (<5 µAU)
- Lamp optimization for the best s/n performance over the lifetime of a Deuterium lamp
- Flexible sampling rates, from 1 to 80 Hz, for normal and fast LC separations
- Independent optimization of sampling rates and filter time constants gives you the ability to tune your detector to your separation goals
- Enhanced thermal wander management reduces baseline wander due to fluctuations caused by ambient conditions
- Flow cell options to match your separation scale and type
- Accessible detector components for easy operation and maintenance
Waters 2489 UV/Vis detector
Condition | Specification |
Wavelength range | 190 to 700 nm |
Bandwidth | <5 nm |
Wavelength accuracy | +1.0 nm |
Wavelength repeatability | +0.1 nm |
Single λ dry noise | <5 µAU (at 230 nm, 1 –second digital filter, 30-second segments, 10-Hz data rate, analytical cell) |
Wet noise | <8 µAU (at 230 nm, 1-second digital filter 30-second segments, 10-Hz data rate, analytical cell) |
Dual channel noise (dry) | <35 µAU (measured at 230 nm, running at 230 to 280 nm with 2-second filter at 1-Hz data rate; dry, analytical cell) |
Dual channel noise (wet) | <45 µAU (measured at 230 nm, running at 230 to 280 nm with 2-second filter at 1-Hz data rate; 1 ml/min, flow of acetonitrile, analytical cell) |
Linearity | <5% at 2.5 AU, propylparaben at 257 nm |
Drift | 1.0 x 10-4 AU/hr. max (∆T = +2°C/hr.) |
Thermal drift | 1.0 x 10-4 AU/°C max (∆T = +2°C/hr.) |
Sensitivity setting range | 0.0001 to 4.0000 AUFS |
Filter setting range | Single-wavelength: 0.1 to 5.0 seconds Dual-wavelength: 1 to 50 seconds |
Time-constant | To 5.0-second Hamming filter |
Digital data rates | 1,2,5,10,20,40,80 Hz (single channel) 1,2 Hz (dual channel) |
Analog output data rates (single λ mode) | 10,20,40,80 Hz (channel A) 10 Hz only (channel B) |
Lamp source | 30-W High Brightness Deuterium lamp, 0.5 nm aperture, pre-aligned 2000 silicon photodiodes (matched pair) |
Second-order filter | Automatic for wavelengths > 370 nm |
Wavelength calibration filter | Erbium filter, used at startup or on demand |
Flow cell | TaperSlit flow cell design |
Nitrogen purge | Purge fitting present on optics bench |
Path length | 10 mm (standard analytical) |
Cell volume | 10 µL (standard analytical) |
Pressure limit | 6895 kPa (69 bar,1000 psi) |
Materials | 316 stainless steel, fused silica, Tygon |
Operating temperature | 4 to 40 °C (39 to 104 °F) |
Operating humidity | 20% to <95%, noncondensing |
Shipping and storage | -30 to +60 °C |
Shipping and storage humidity | 0% to <95% noncondensing |
This refurbished Waters 2489 UV/Visible Detector is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.