Wavelength range | 190 to 800 nm |
Light source | Prealigned, Deuterium lamp |
Spectral Resolution/Optical Bandpass | 1.2 nm per photodiode with a total of 512 photodiodes, digital and optical (3D-mode) |
Digital Resolution | 1.2 nm – 600 nm (2D mode) |
Wavelength accuracy | ±1 nm |
Linearity range | 5% at 2.0 AU, propylparaben, at 256 nm |
Spectral resolution | 1.2 nm |
Baseline noise | ±1.5 × 10–5 AU peak-to-peak, dry, at 254 nm |
Drift | 1 × 10–3 AU/hour at 254 nm (after warm-up) ∆T ≤ 1°C per hour |
Operation temperature | 4 to 40 °C (39 to 104 °F), not to vary more than 1 °C per hour (to prevent drift) |
Relative humidity | 20 to 80%, noncondensing |
Power | Grounded ac, 100 to 240 Vac |
Power requirements | 100-230 VAC |
Frequency range | 50 to 60 Hz |
Path Length | 10 mm (standard analytical cell) |
Cell volume | 8 µl (standard analytical cell) |
Pressure Limit | 1000 psi (standard analytical cell) |
Wetted materials | 316 stainless steel, fused silica Tefzel |
Dimensions | Width: 11.5 in. (29 cm) Depth: 24 in. (61 cm) Height: 8.5 in. (22 cm) |
Weight | 31.5 lbs (14.3 kg) |
Injection volume range |
|
Accuracy | +0.2 µL, measured by fluid weight removed from vial with 10.0 µL injections averaged over 20 injections using standard 100-µL syringe |
Injection Linearity | >0.999 (See ACQUITY UPLC I-Class system specification document for conditions) |
Maximum sample capacity | Any two of the following:
|
Sample compartment temperature range |
|
Temperature accuracy | +0.5°C at sensor |
Temperature stability | +1.0°C at sensor |
Injection needle wash | Integrated, active, programmable |
Minimum sample required | 3µL residual, using Waters’ total recovery 2-mL vials |
Carryover (UV) | <0.004% <0.001% typical |
Advanced sample manager capabilities | Auto-dilution and auto-addition |
Primary wetted materials | 316L stainless steel, polyimide, PEEK blend, DLC, PPS |
Number of solvents | Up to four, in combination of two, A1 or A2 and B1 or B2 |
Solvent conditioning | Integrated vacuum degassing, six lines with two allocated for the injector needle wash/purge solvents |
Gradient formation | High pressure mixing, binary gradient |
Gradient profiles | 11 gradient curve [ including linear, step (2), concave (4), and convex (4)] |
Primary check valves | Intelligent Intake Valves |
Flow accuracy | +1.0% of set flow rate at 0.500 mL/min, as per SystemsQT |
Flow precision | 0.075% RSD or 0.01 min SD, (0.2 to 2.0 mL/min), whichever is greater using premixed solvent |
Composition ripple (baseline noise) | <1.0 mAu |
Pump compositional precision | <0.2% RSD, or 0.02 min SD, whichever is greater (from 0.2 to 2.0 mL/min) |
Pump compositional accuracy | +0.5% absolute from 5% to 95%, 0.2 to 2.0 mL/min |
Compressibility compensation | Automatic, no user intervention required |
Priming | Wet priming runs at a flow rate of 4 mL/min |
Pump seal wash | Equipped with a programmable active wash system to flush the rear of the high pressure seals and the plungers |
Flow ramping | Automatic |
Primary wetted materials |
316L stainless steel, UHMWPE blend, MP35N, titanium alloy, gold, sapphire, ruby, zirconia, Nitronic 60, DLC, fluoropolymer, PEEK and PEEK blend |
Mixing options |
|
This Waters UPLC System includes:
- Waters 2996 PDA Detector
- Waters Sample Manager
- Waters Binary Solvent manager
This refurbished Waters UPLC System is in great condition. It has been tested at SpectraLab Scientific. A 90 day warranty is included.