Item | Specification |
Wavelength range | 190 to 800 nm |
Light source | Prealigned, Deuterium lamp |
Spectral Resolution/Optical Bandpass | 1.2 nm per photodiode with a total of 512 photodiodes, digital and optical (3D-mode) |
Digital Resolution | 1.2 nm – 600 nm (2D mode) |
Wavelength accuracy | ±1 nm |
Linearity range | 5% at 2.0 AU, propylparaben, at 256 nm |
Spectral resolution | 1.2 nm |
Baseline noise | ±1.5 × 10–5 AU peak-to-peak, dry, at 254 nm |
Drift | 1 × 10–3 AU/hour at 254 nm (after warm-up) ∆T ≤ 1°C per hour |
Operation temperature | 4 to 40 °C (39 to 104 °F), not to vary more than 1 °C per hour (to prevent drift) |
Relative humidity | 20 to 80%, noncondensing |
Power | Grounded ac, 100 to 240 Vac |
Power requirements | 100-230 VAC |
Frequency range | 50 to 60 Hz |
Path Length | 10 mm (standard analytical cell) |
Cell volume | 8 µl (standard analytical cell) |
Pressure Limit | 1000 psi (standard analytical cell) |
Wetted materials | 316 stainless steel, fused silica Tefzel |
Dimensions | Width: 11.5 in. (29 cm) Depth: 24 in. (61 cm) Height: 8.5 in. (22 cm) |
Weight | 31.5 lbs (14.3 kg) |
This Refurbished Waters NanoACQUITY UPLC System with Waters ACQUITY 2996 PDA comes with the following:
- Waters Acquity 2996 PDA with Nano scale flow cell installed
- Waters NanoACQUITY BSM Binary Solvent Manager
- Waters NanoACQUITY ASM Auxiliary Solvent Manager
- Waters NanoACQUITY SM Sample Manager with attached column oven
- Waters Sample Tray
- PC with Suitable software and all essential cables required for full operation
We encourage you to contact us to speak with one of our expert and professional team.