Type |
Specification |
Injection modes |
Partial loop filling Complete loop filling |
Injection range |
0.1 – 5000 µL in 0.1 µL increments. Recommended injection range 10 – 5000 µL |
Loop size |
Max. loop size 5 ml (selectable range: 5 µL – 50000 µL) |
Area precision Complete loop fill |
Typically < 0.5% RSD (50 µL loop, overfill factor 3 or greater) |
Area precision Partial loop fill |
Typically < 1.0% RSD (5000 µL loop, 1000 µL injection volume, draw/eject speed 20 mL/min) |
Sample viscosity range |
0.2 – 5 cp |
Sample capacity |
2 well-plates (96 and 384) plus 10 additional 2-mL vials 108 x 2-mL vials in 2 x 54 vial plate plus 10 additional 2-mL vials 30 x 6-mL vials in 2 x 15 vial plate plus 10 additional 2-mL vials 54 Eppendorf tubes (0.5/1.5/2.0 mL) in 2 x 27 Eppendorf tube plate Vial septa and closing mats must be pre-slit. |
Carryover |
< 0.1% with rinse factor > 3 (see test method listing in manual) |