Specification |
|
Source | D2 + W lamps, Hg lamp for wavelength calibration |
Wavelength range | 190-900 nm |
Noise | <0.25 x 10-5 AU at 250 nm and 600 nm |
Drift | <0.5 x 10-3 AU/h at 250 nm and 600 nm |
Slit width | 1 nm, 4 nm |
Response time constants | 0.01/0.02/0.05/0.1/0.5/1/2 s |
Data collection rates | Shortest data interval 10 ms or 100 Hz |
Standard cell | 5 mm/ 3.2 µL |
FDA/GXP functions | Automatic wavelength calibration with various D2 and Hg spectral lines Maintenance logbook: data of last lamp replacement (D2, W and Hg lamps), lamp energy, lamp usage, number of ignitions In conjunction with the EZChrom Elite Software: Documentation of instrument serial number and configuration Detailed audit trails and logbooks |
This refurbished Hitachi LaChrom ULTRA L-2455U Diode Array Detector is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.