Specifications
Source | Xenon-lamp (150 W), Hg lamp for wavelength calibration |
Wavelength range | Excitationb: 200-850 nm, emission: 250-900 nm (at wavelengths > 731 nm another photomultiplier has to be used) |
Spectral bandwidth |
Excitation: 15 nm Emission: 15 nm or 30 nm, adjustable |
Sensitivity | S/N for the Raman peak of water: > 550 (baseline method, emission slit width 30 nm) for vthe high-speed cell, > 900 for the standard cell (option) |
Response time constants | 0.01/0.02/0.05/0.1/0.5/1/2 s |
Data collection rates | Shortest data interval 10 ms or 100 Hz |
Programmes |
Programmable parameters:
|
Flow-through cell | 3 µL (ultra-high-speed cell) |
FDA/GXP functions |
In conjunction with the EZChrom Elite software:
|
Safety | Leak sensor, error handling |
Power requirements | 1oo-240 V (U50/60 Hz), 260 W |
Dimensions | 340 (W) x 400 (D) x 300 (H) mm |
This Hitachi LaChrom ULTRA L-2485U FL Detector is in great condition. It has been tested here at SpectraLab Scientific.
Includes a 90 day warranty