Specifications
Heated Devices
- Five heated zones standard: Two detectors, Two inlets, one auxiliary
- Methods stored: Two
Split/Splitless Capillary
- Range to 400°C in 1°C increments
- Back-pressure design permits independent adjustment of split flow rate without affecting column flow
- Septum purge builtin at 3 ml/min
- Accepts columns up to 1.2-mm od
- Multi mode design includes split and splitless injection
- Splitless purge time variable in 0.01-min elements
Thermal Conductivity Detector
- Range to 400°C
- Single-filament(single column) design has fluidic switching of reference and analytical carrier flows
- Minimum detectable:400 pg/ml carrier — equivalent to 1 pmm neon in 1 ml air
- Linear dynamic range:+/- 5% over 105 range
- Digital gain setting time-programmable through the keyboard Conditions: detector
- -100°C, 45 ml/min switching and 30 ml/min analytical flow of helium, propane sample
Flame Photometric Detector
- Range to 300°C
- 100-530 u id capillary columns fit through the transfer line to the base of the flame
- Two combustion modes: air/hydrogen or oxygen/hydrogen
- Detector signal display to view when flame is lit
- Push-button flame ignition
- Sulphur mode:Sensitivity to 20 pg S/second, Range 103, Selectivity 105 g S/g C
- o Phosphorus mode:Sensitivity to 0.9 pg P/second, Range 104, Selectivity 106 g P/g C
Column Oven
- Operating range: 4°C above ambient to 450°C
- Set point entry: 1°C for temperatures, 0.1°C for program rates
- Programming: Rates 0.1 to 70°C per min, 650 min maximum run time, three ramps with initial/final holds
This refurbished HP 5890 II GC with Dual TCDs, Dual EPC controlled on-column Inlets is in great condition. It has been tested at SpectraLab Scientific.
Installation and extended warranty available upon request.
A 90 day warranty is included.